HFG (High Frequency Generator) For C-Arm X-ray, Mobile |
|
|
|
MODEL : JWE-J24
Features | High Power Density
High efficiency Inverter Power
Microprocessor controller
Input, PFC Status Monitor
Protections: Over Voltage / Over Load / Thermal
Power Factor Circuit Design |
|
|
|
|
OUTPUT | * OUTPUT POWER(C-Arm) | 3.2KW |
KV Range | 40 ~ 125KV ¡¾ 5%+1KV |
mA Range (Fluoroscopy) | 0.1 ~ 5mA ¡¾ 5%+1mA |
mA Range (Boost) | 0.2 ~ 10mA ¡¾ 5%+1mA |
mA Range (Radiography) | 25 ~ 40mA ¡¾ 5%+1mA |
Output Time (Fluoroscopy) | 0 ~ 5 min |
mAs Range (Radiography) | 0.4mAs ~ 240mAs |
*Output Power(Mobile) | 2.4KW |
INPUT | VOLTAGE RANGE | 176AC ~ 264VAC |
FREQUENCY RANGE | 47 ~ 63Hz |
POWER FACTOR | PF ¡Ã 0.9 |
INRUSH CURRENT | < 50A / 220VAC |
LEAKAGE CURRENT | < 3.5mA / 230VAC |
PROTECTION | HV Over Voltage | 110% Over Shut Down |
HV Over Current | 110% Over Shut Down |
Thermal | HV Tank 65¡É ¡¾ 5¡É |
ENVIRONMENT | WORKING TEMP. | -10 ~ +50¡É |
WORKING HUMIDITY | 10 ~ 90% RH |
STORAGE TEMP., HUMIDITY | -20 ~ +60¡É |
SAFETY & EMC | WITHSTAND VOLTAGE | IN-OUT, IN-Case : 1.5KVAC |
ISOLATION RESISTANCE | IN-OUT, IN-Case : 100M OHMS / 500VDC |
SURGE | Line-Line : 1KV, Line-GND : 2KV |
OTHERS | MTBF | 200Khrs min.(25¡É) |
| |
|
|
|
|
|
|
|
|
|