HomeÁ¦Ç°¾È³»Á¦Ç°¾È³»
HFG (High Frequency Generator) For C-Arm X-ray, Mobile
 
  MODEL : JWE-J24
Features
High Power Density
High efficiency Inverter Power
Microprocessor controller
Input, PFC Status Monitor
Protections: Over Voltage / Over Load / Thermal
Power Factor Circuit Design
 
MODEL
JWE-J24

OUTPUT

* OUTPUT POWER(C-Arm)

3.2KW

KV Range

40 ~ 125KV ¡¾ 5%+1KV

mA Range (Fluoroscopy)

0.1 ~ 5mA ¡¾ 5%+1mA

mA Range (Boost)

0.2 ~ 10mA ¡¾ 5%+1mA

mA Range (Radiography)

25 ~ 40mA ¡¾ 5%+1mA

Output Time (Fluoroscopy)

0 ~ 5 min

mAs Range (Radiography)

0.4mAs ~ 240mAs

*Output Power(Mobile)

2.4KW

INPUT

VOLTAGE RANGE

176AC ~ 264VAC

FREQUENCY RANGE

47 ~ 63Hz

POWER FACTOR

PF ¡Ã 0.9

INRUSH CURRENT

< 50A / 220VAC

LEAKAGE CURRENT

< 3.5mA / 230VAC

PROTECTION

HV Over Voltage

110% Over Shut Down

HV Over Current

110% Over Shut Down

Thermal

HV Tank 65¡É ¡¾ 5¡É

ENVIRONMENT

WORKING TEMP.

-10 ~ +50¡É

WORKING HUMIDITY

10 ~ 90% RH

STORAGE TEMP., HUMIDITY

-20 ~ +60¡É

SAFETY & EMC

WITHSTAND VOLTAGE

IN-OUT, IN-Case : 1.5KVAC

ISOLATION RESISTANCE

IN-OUT, IN-Case : 100M OHMS / 500VDC

SURGE

Line-Line : 1KV, Line-GND : 2KV

OTHERS

MTBF

200Khrs min.(25¡É)